Herunterladen Inhalt Inhalt Diese Seite drucken

Operating Elements And Cable Connections; Connection For 5-Pole Connecting Cable; Mounting; Emissivity - KLEIBER KG 740 - LO Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5 Technical Description

5.3 Operating elements and cable connections

The operating elements and cable connections are located on the rear side of the device.
Push button
Potentiometer
BNC output
pilot light

emissivity

Light spot
Emissivity
0,6
Output 0 - 10 V
0,7
0,8
0,9
1
Type:
KGA 740 - LO
No.:
35123
Measuring range:
300
1400
Input:
24 V AC / DC
Output:
4 - 20 mA
5-pin plug connection
Figure 6: Rear side pyrometer

5.4 Connection for 5-pole connecting cable

The 5-pole plug to connect the instrument is at the rear side of the unit. The contacts of the 5-pole plug are
arranged as follows:
4
3
2
5
1
Figure 7: Connection 5-pole ca-
ble

5.5 Mounting

To mount the pyrometer, a threaded drilling M12 is located on the underside of the unit. The pyrometer can be
attached to different mountings.
We recommend to use the retaining bolt supplied with the pyrometer or the optional available ball and socket
mounting support with clamp or thread. This mounting plate ensures secure assembly of the pyrometer (for
order data see Accessories (section 3.3) at page 5).
5.6 Emissivity
The emissivity is the relationship between infrared energy radiated from an object and the radiation energy of
a perfect emitter (black emitter) at the same temperature and the same spectral range.
The emissivity is material-dependent and of a considerable size in order to be able to determine the temper-
ature of an object accurately without contact. The emissivity of the object being measured must therefore be
known and be adjusted at the pyrometer.
Typical emissivities for various materials are available on our homepage at www.kleiberinfrared.com or in rele-
vant literature.
The instrument is calibrated at black body radiation ( = 1,00). Real objects have an emissivity less than
1.00. This difference can be adjusted with changing the emissivity value from the pyrometer.
12
Element
Push button for pilot light
(Alternative: Instead through
lens sighting)
0,5
0,4
BNC output
0,3
0,2
Potentiometer emissivity
0,1
°C
5-pin plug connection
Potentiometer can be set with the help of a suitable screw driver without
any violence.
Contact
1
2
3
4
5
Meaning
Switch on/ off the pilot light
Measurement output
0 . . . 10 V
Adjustment emissivity
0.1 . . . 1.0
Power supply connection and current
output
Color cable
Meaning
core
White
+24 V AC / DC supply voltage
Brown
0 V AC / DC supply voltage
Green
Ground output
Yellow
Analog output 0 / 4 . . . 20 mA
Green-Yellow
PE
=
KS / KG / KGA 740 - LO

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis