5 Operation

5.2 Operation
Before operating the KLEIBER KG 315 - K USB the looking though lens is to be adjusted at a bright, white
background (e.g. a sheet of paper) by turning the ocular until the circle-shaped measuring field is visible
sharply at the relaxed eye.
We recommend the using of the grey filter by screwing in the thread of the ocular at high measuring tempera-
tures.
Focusing is done by the looking through lens. To adjust the distance loosen the clamping ring and move the
tube by turning until the measured object is focused sharply. Tighten the clamping ring after focusing.
The measuring button is a two step button. By pressing and holding the button in the first position, the infrared
thermometer will be activated and switched to the measuring mode. The point to be measured can be searched
by moving the instrument. Depending on the chosen configuration the measured value, the average value or
the peak value are displayed in the looking through lens and at the rear panel display.
By pressing the measuring button to the second step the value will be stored to the next storage position. The
success of the writing procedure is shown by a long beep. If there are any problems writing the value (e.g.
exceeding limits) ten short beeps are to be heard. The measured values are stored according to the used
modes. For the Classic mode the measured values are stored one after another, that means one row will be
filled up until one hundred measuring values. After the row is filled the instrument starts to fill the next row from
point 001.

5.2.1 Recording modes

The KLEIBER KG 315 - K USB can be equipped with various measuring modes. Which it is or they are, is
listed on the inspection sheet.
Classic
Standard mode. There can 2000 measuring values in 20 rows at 100 measuring values
be saved.
Classic Extended
Standard mode with additional options. The number of rows and values is variable. In
addition, more settings are available in the configuration menu.
Ways
Ways are measurement points with definable names in any order. The pyrometer can
permanently store, in dependence on its size, several routes. These can be processed
and the measurement values are stored by the individual measuring points.

5.2.2 Internal rechargeable battery charge

Insert the rechargeable 1.2 V batteries, come with the instrument, into the Battery compartment of the infrared
thermometer.
CAUTION:
Pay attention to polarity!
Connect the recharger type Charging unit ACS 110 for internal charging with the recharging joint on the front
side of the instrument. Recharging starts after a short testing phase (app. 10 sec.). After recharging the red
LED switches off and the trickle charging will start automatically (green LED glows). If a defective recharge-
able battery is inserted, the red LED will flicker. A defective rechargeable battery is easy to locate by using the
optional recharger Charging unit Voltcraft for external charging and must be replaced.
CAUTION:
Never connect the recharger to the instrument with no rechargeable batteries inserted!
8