Alternating current

Range
40.00 A
400.0 A
Frequency range: 50 - 60 Hz; Overload protection: 600 V, 400 A
*Measuring position error: Accuracy deviation for non-centred measurement position: +1%
*Accuracy: 40 - 60% of the measurement range; +(1.5% ± 10) outside of this range
TrueRMS crest factor (CF) for non-sinusoidal signals: max. 3.0
CF >1.4 - 2.0
+ 1%
CF >2.0 - 2.5
+ 2.5%
CF >2.5 - 3.0
+ 4%
Direct current (VC-523 only)
Range
40.00 A
400.0 A
Overload protection 600 V, 400 A
*Measuring position error: Accuracy deviation for non-centred measurement position: +1%
*Accuracy: 40 - 60% of the measurement range; +(1.5% ± 10) outside of this range
AC voltage
Range
4.000 V
40.00 V
400.0 V
600 V
Frequency range: 50 - 100 Hz; Overload protection: 600 V; Impedance: 10 MΩ
*Accuracy: 40 - 60% of the measurement range; +1% outside of this range
TrueRMS crest factor (CF) for non-sinusoidal signals: max. 3.0
CF >1.4 - 2.0
+ 1%
CF >2.0 - 2.5
+ 2.5%
CF >2.5 - 3.0
+ 4%
54
Resolution
0.01 A
0.1 A
Resolution
0.01 A
0.1 A
Resolution
0.001 V
0.01 V
0.1 V
1 V
Accuracy*
±(2% + 12)
±(2.8% + 8)
Accuracy*
±(2.8% + 12)
±(2.8% + 8)
Accuracy*
±(1.5% + 7)