Herunterladen Inhalt Inhalt Diese Seite drucken
Vorschau ausblenden Andere Handbücher für MN93:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 42
Current clamps, MiniFlex
the current flowing in a cable without opening the circuit. They also insulate the user
from dangerous voltages in the circuit.
The choice of current sensor to be used depends on the current to be measured
and the diameter of the cables.
Measurement range *
Clamping diameter
Measurement range *
Clamping diameter
Measurement range *
Clamping diameter
Measurement range *
Clamping diameter
Measurement range *
Clamping diameter
* : The measurement ranges are those of the sensors. In some cases, they may differ
from the ranges that can be measured by the analyser with which they are associated.
„ For three-phase measurements, start by marking the accessories using the col-
oured rings supplied with the device: a different colour for each current terminal.
„ Connect the current sensors to the current terminals of the device.

2. USE

and AmpFlex
®
®
MN93
0.05 - 240 A
20 mm
C193
1 - 1.200 A
52 mm
MiniFlex
®
MA193
250 mm
100 mA - 10 000 A
70 mm
MiniFlex
®
MA194
250 mm
100 mA - 10 000 A
70 mm
AmpFlex
®
A193
450 mm
100 mA - 10 000 A
140 mm
flexible current sensors are used to measure
MN93A
5 A
0.005 - 6 A
20 mm
MiniFlex
MA193
350 mm
100 mA - 10 000 A
100 mm
MiniFlex
MA194
350 mm
100 mA - 10 000 A
100 mm
AmpFlex
A196A
610 mm
100 mA - 10 000 A
190 mm
34
MN93A
100 A
0.1 - 120 A
20 mm
PAC93
1 - 1.200 A
39 mm
MiniFlex
®
MA196
350 mm
100 mA - 10 000 A
100 mm
MiniFlex
®
MA194
1000 mm
100 mA - 10 000 A
320 mm
AmpFlex
®
A193
800 mm
100 mA - 10 000 A
250 mm
®
®
®

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis