4 Standard configuration
CO
Measuring Range
0 to 500 ppm 0 to 100 ppm
Vibrating alarm
Yes
1)
Alarm Threshold A1
30 ppm
acknowledgeable
Yes
latching
No
1)
Alarm Threshold A2
60 ppm
acknowledgeable
No
latching
Yes
3)
Fresh air calibration
on
Bump Test Mode
off
Life Signal
off
Turning device off
always
1) Observe the country-specific settings. The alarm thresholds must be adjusted according to national regula-
tions, if necessary.
2) For O
A1 is the lower alarm threshold, used to indicate oxygen deficiency.
2
3) User can select to fresh air calibrate after startup.

5 Operation

The device may only be used in areas subject to explosion hazards which are
explicitly covered under the Ex Approvals which have been given to this device.
Electrical operating safety (Ex protection) is not guaranteed in an oxygen-
enriched atmosphere.
*) Please be aware of special settings by customer request.
*)
H
S
O
2
2
0 to 25 vol.-%
Yes
Yes
5 ppm
19 vol.-%
Yes
No
No
Yes
10 ppm
23 vol.-%
No
No
Yes
Yes
on
on
off
off
off
off
always
always
WARNING
Whenever the Pac 3500/5500 is opened, a bump test and/or a calibration
must be performed afterwards. This includes each battery replacement as
well as each sensor replacement in the Pac 3500/5500. Non-compliance
will mean that the proper functioning of the device cannot be guaranteed
which may impair the accuracy of the measurements.
2)
Before making safety-related measurements, check the adjustment and
adjust as needed, and check all alarm elements. If there are national regula-
tions, the bump test must be performed according to these regulations.
Faulty adjustment can lead to incorrect measurement results and, therefore,
also cause severe health risks.
Remaining life of the device (only valid for Pac 3500)
– Once activated check the remaining life by pressing [+] while device is
turned off. The remaining time in days will be shown. After another press of
[+] "d" will be shown.
After another press of [+] the gas to be measured will be shown, e. g. "750",
"d", "CO".
Activating a new device
– Press and hold [+] for approx. 3 seconds while "3, 2, 1" appears in the dis-
play. The device's usable life is now started. The gas to be measured will
be shown. After 10 seconds the display will turn off or if [+] is pressed again,
"d" will be shown. After 10 seconds the display will turn off or after another
press of [+] the remaining time in days will be shown. After 10 seconds the
display will turn off.
5.1 Turning the device on
– Press and hold [OK]. The display counts down until startup: "3, 2, 1".
All display segments are lit. Next, the LED, Alarm and Vibrating alarm are
activated in sequence. Check these before each use.
– The device will perform a self test.
– The software version and the gas name are displayed.
– The number of days of remaining operation are shown, e. g. "750", "d" (only
valid for Dräger Pac).
– The A1 and A2 alarm limits are displayed.
– If the calibration interval function is activated the days remaining until the
next calibration, e.g., » CAL « then » 20 « are displayed.
– If the bump test interval function is activated, the time until the bump test in-
terval elapses is displayed in days, e.g., » bt « then » 123 «.
– The first warmup time in seconds is displayed alternating with the letters
"SEC".