Proceed as follows for fuse replacement:
- Disconnect all measuring leads from the meter and switch it off.
- Open the battery compartment as described in the chapter "Inserting and Changing the Batteries".
- The fuses can be accessed now.
- Replace the defective fuse(s) with a new fuse of the same type and nominal voltage. The fuses have
the following values:
Fuse
Measuring range Current µA/mA
Characteristic
Dimensions
Type example
- Close the housing again carefully by proceeding in reverse order.
Using mended fuses or bridging the fuse holder is not admissible for safety reasons.
This can lead to fires or to arc explosions. Never operate the measuring instrument
when it is open.

Disposal

Old electronic devices are hazardous waste and should not be disposed of in the household
waste. When the device has become unusable, dispose of it in accordance with the current
statutory regulations at the communal collection points. Disposal in the domestic waste is not
permitted.

Disposal of used batteries!

As a consumer you are required (Battery Ordinance)to responsibly dispose of all used batteries and
rechargeable batteries; it is forbidden to throw them away with the normal household waste!
Batteries/rechargeable batteries containing harmful substances are marked with the follow-
ing symbols, they point out that they are not allowed to be disposed of in the domestic
refuse. The symbols for dangerous heavy metal constituents are: Cd = cadmium, Hg =
mercury, Pb = lead. You can return used batteries / storage batteries free of charge to any
collecting point of your local authority, to our stores or to any store where batteries / storage
batteries are sold.
You thus fulfil the legal requirements and make your contribution to the protection of the envi-
ronment!
40
F1
FF 500 mA/660V
5 x 20 mm
SIBA® Fuse
L-no. 70 180 40
F2
400 mΩ, 4/40Ω
FF 500 mA/660V
5 x 20 mm
SIBA® Fuse
L-no. 70 180 40