Remote with sensor terminal housing
● SITRANS FCS400 sensor
● SITRANS FCT030 transmitter with terminal
housing assembled
● Sensor cable
● Packet of cable glands
● Quick Start guide
● CD containing software, certificates and device
manuals
Note
Scope of delivery may vary, depending on version and add-ons. The contents list in the product package lists all included
parts.
Note
Transmitter installation
The installation of the FCT030 transmitter is described in the SITRANS FCT030 Quick Start (A5E03650800).
Sensor overview
Lid-lock
Cable feed-through (M12 socket or gland)
Sensor front end (DSL) (Remote configuration only)
Plug and threaded port for for example pressure guard
Sensor enclosure
Process connections
Location in the system
The optimum location in the system depends on the application:
● Liquid applications
Gas or vapor bubbles in the fluid may result in erroneous measurements, particularly in the density measurement.
– Do not install the flowmeter at the highest point in the system, where bubbles will be trapped.
– Install the flowmeter in low pipeline sections, at the bottom of a U-section in the pipeline.