3.2
Transport and storage
Permissible ambient temperature range during transport and storage:
• Cartridge not filled with water (state of delivery): -10 °C / 14°F up to +40 °C / 104°F.
• Cartridge filled with water: +4 °C / 39.2 °F up to +40 °C / 104°F.
The cartridge must not freeze since it was filled with water the first time.
Danger of frost.
Damage to the cartridge.
∅ Do NOT expose the cartridge to temperatures below +4 °C / 39.2 °F.
Permissible ambient humidity: 85 % r.H. max., non-condensing
Intermediate storage in a closed and dry room.
3.3
Installation and ambient conditions
Set up BINDER Pure Aqua Service upright on an even surface. The connected cartridge is a pressurized
device. The content may irritate eyes and skin.
Opening by force or damage to the cartridge.
Risk of injury.
Irritation of skin and eyes.
Damage to the cartridge and its environment.
∅ Do NOT open.
Permissible ambient temperature range during operation: +4 °C / 39.2 °F up to +40 °C / 104°F (climatic
chamber: +18 °C / 64.4°F to +32 °C / 89.6°F.)
Permissible ambient humidity: 85 % r.H. max., non-condensing.
To avoid any possible water damage, provide a floor drain at the location of the device. Select a suitable
installation site to avoid any consequential damage by splashing water.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power supply unit is easily accessible and can be easily disconnected in case of
danger.
The unit must not be installed and operated in potentially explosive areas.
Explosion hazard.
Danger of death.
∅ Do NOT operate the unit in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.
Mounting instructions BINDER Pure Aqua Service (E3) 12/2017
CAUTION
CAUTION
WARNING
page 28/57