Note to the surgeon or surgeons involved: Prior to the first implantation of the Atrostim PNS, please
contact the manufacturer or the dealer to discuss about possible surgical assistance or training.
1 COMPONENTS OF ATROSTIM PNS
Most of the components, which are implanted into the patient, are covered with
silicone rubber medical grade. Silicone rubber is extremely resistant to bending,
resistant to stretching, but extremely vulnerable by sharp instruments. These
components should never come into contact with surgical forceps, tweezers or other
instruments with cutting blades or edges. Cutting may cause leakage of body fluids
into the components, which might cause corrosion of wires and electrodes and
malfunction of electronic components. Cuts that may cause leakage may be too small
to be recognized with the naked eye.
The following components of Atrostim Phrenic Nerve Stimulator (PNS) system are
in contact with or are implanted into the patient (Fig. 1) (one side shown only):
Energy transfer
coil
Fig. 1. Components of Atrostim PNS, which are in contact with or are implanted
into the patient.
April 2008
Surgical procedure
for pediatric patients
Electrode
Compound
connector
Implant
stimulator
LEFT CH.
RIGHT CH.
FULL BATTE RY
LOW BA T T ERY
MODE
RIGHT
LEFT
BILAT
V OLUME
LOW
NORM AL
HIGH
ADJUST MENT
SET
SELECT
SIGH
RATE
Stimulus
BATT
LOCK
ON
OFF
TEST
controller
Surgical procedure for pediatric patients
Extension
lead
1
Respiratory
center
Spinal cord
C
- C
1
4
Cell bodies
of phrenic
nerve
C
- C
(C
)
3
5
6
Phrenic nerve
Motor end plate
Diaphragm
muscle