12 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
The displayed weight does not glow.
The displayed weight is permanently
changing
The weighing value is obviously
wrong
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
PCB-BA-e-1718
Possible cause
• The balance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
• Power supply interrupted.
Batteries are inserted incorrectly or empty
• No batteries inserted.
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with
foreign matter.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
41