Herunterladen Inhalt Inhalt Diese Seite drucken

KERN PCB 200-2 Betriebsanleitung Seite 83

Präzisionswaage
Vorschau ausblenden Andere Handbücher für PCB 200-2:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

rE Cr:
The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.
Take into account that the following remote control commands must be sent
without a subsequent CR LF.
s
Function:
w
Function:
t
Function:
h. Format for stable values for weight/quantity/percentage
1
2
3
4
M
S
N
N
1
2
i. Format in case of fault
1
2
3
4
B
B
B
B
j. Format for unstable values for weight/quantity/percentage
1
2
3
4
M
S
N
N
1
2
Symbols
M
Blank or M
S
Blank or minus sign (-)
N
... N
10 numeric ASCII codes for weight values including decimal places or
1
10
blanks
U
... U
3 ASCII codes for weighing unit pcs. / % / or blank
1
3
B
Blank
E, o, r
ASCII code or "E, o, r"
CR
Carriage Return
LF
Line Feed
PCB-BA-e-1718
Stable weighing value for the weight is sent via the RS232
interface
Weighing value for the weight (stable or unstable) is sent
via the RS232 interface
No data are sent, the balance carries out the tare function.
5
6
7
8
9
N
N
N
N
N
3
4
5
6
7
5
6
7
8
9
B
B
B
B
B
5
6
7
8
9
N
N
N
N
N
3
4
5
6
7
10 11 12
13 14 15 16 17
N
N
N
B
U
8
9
10
10 11 12
13 14 15 16 17
B
B
E
r
r
10 11 12
13 14 15 16 17
N
N
N
B
B
8
9
10
18
U
U
CR LF
1
2
3
18
o
r
CR LF
18
B
B
CR LF
39

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Pcb 100-3Pcb 250-3Pcb 350-3

Inhaltsverzeichnis