 Press
repeatedly until LinEAr is
displayed
 Ensure that there are no objects on the
weighing pan.
 Start linearisation with
first adjustment weight will be displayed.
 Place adjustment weight and acknowledge by
. The scales will change to zero display.
 Take away adjustment weight. After a short
time the value of the second adjustment weight
appears in the display.
 Place second adjustment weight and
acknowledge by
to zero display.
 Take away adjustment weight. After a short
time the value of the third adjustment weight
appears in the display.
 Place third adjustment weight and
acknowledge by
to zero display.
 Take away adjustment weight. After a short
time the value of the forth adjustment weight
appears in the display.
 Place forth adjustment weight and
acknowledge by
to zero display.
 Take away adjustment weight. After a short
time the value of the fifth adjustment weight
appears in the display.