BYK wave-scan dual Betriebsanleitung Seite 48

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Testing the instrument
14
Due to the underlying measurement
principle, no calibration of the
instrument is required. It is
recommended, however, to check the
functionality of the instrument at
regular intervals (about once every 3
months). The reference tile included
with delivery is provided for this
purpose.
Select the "Standard" in the Measure
menu. The configuration of this
organizer is fixed and cannot be
changed.
The unit then switches to the
measurement mode.
Place the instrument on the test tile
marked „Standard" or „Standard 1" as
shown above and perform three
measurements.
wave-scan dual only:
After having finished these readings,
you are prompted to take three more
readings on Standard 2.
Place the instrument on the test tile
marked Standard 2 and perform three
measurements.