6. Notes for mounting and operating in hazardous areas
Ambient temperature range
Model T24.10.**2
GB
II 1G EEx ia
Model T24.10.**9
II 3G EEx nL/nA
6.3 Safety-engineering values
6.3.1 Power supply / 4 ... 20 mA-loop
The following safety-relevant maximum values must not be
exceeded:
Model T24.10.**2
II 1G EEx ia
Model T24.10.**9
II 3G EEx nL/nA
Following have an outward effect at the connection terminals j und
i of the transmitter:
effective internal capacitance C
effective internal inductance
12
WIKA Operating instructions temperature transmitter Model T24.10
T4 : - 50 °C ≤ T
≤ + 85 °C
a
T5 : - 50 °C ≤ T
≤ + 75 °C
a
T6 : - 50 °C ≤ T
≤ + 60 °C
a
T4 : - 50 °C ≤ T
≤ + 85 °C
a
T5 : - 50 °C ≤ T
≤ + 65 °C
a
T6 : - 50 °C ≤ T
≤ + 50 °C
a
Voltage:
U
= DC 30 V
i
I
Current:
= 120 mA
i
Power:
P
= 800 mW
i
Voltage:
U
= DC 36 V
i
= 6.2 nF
i
L
= 110 µ H
i