Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ENGLISH
16
Problem
The Philips BlueControl
becomes warm during use.
Model
Rated voltage (V)
Rated frequency (Hz)
Rated input (W)
Class
Optical output classification
Maximum output of optical peak power density
Max. daily dose (J/cm²)
Battery type
Model
Input
Output
Operating conditions
Temperature
Relative humidity
Possible cause
The standard USB plug is
not inserted or inserted
properly into the adapter.
The wall socket is not live.
This is normal.

Specifications

Philips BlueControl is certified as medical non-laser light source equipment.
While in operation it generates Blue Light with the following specification:
PSD1111
100-240
50-60
15W
Medical device Class IIa
Risk Group 1 according EN 60601-2-57, meaning no
risk in normal intended use.
200mW/cm² during pulsed operation.
90
3.7V /rated capacity 5.77 Whr Lithium-ion battery
Solution
Insert the standard USB plug into the
adapter properly.
Connect another appliance to the same
wall socket to check if the wall socket is
live. If the wall socket is live, but the Philips
BlueControl does not charge, contact
Philips.
No action required.
PR3089
100-240V AC / 50-60Hz
5V DC / 5W
from +10°C to +35°C
from 30% to 90%

Werbung