Information for users
A brief overview of the various instrument probes:
Dermatology, gynaecology, surgery
A
recommendation
of which type of instrument
probe is suitable for which area of use can be find in
the accessory list of recommended indications.
Dentistry
Installation of the PSKP syringe
q Before use, fill the PSKP syringe (E4) with air. To do
this, hold the PSKP syringe firmly with one hand and
with the other hand pull the lower section with the
metal end (E4a) downwards. The air inside the trans-
parent air chamber is later converted into cold plasma.
w The NaviTips (E4c) or
Capillary
(E4b)
Tips
must be pushed onto
the PSKP syringe (E4)
before the PSKP probe is
plugged into the con-
verter.
Plug the PSKP syringe with the metal end (E4a) into
the converter (D3).
n
Now switch on the energy control unit by press-
ing the
ON
button until the POWER LED display
illuminates green.
Via the ON button you can reach level 5.
n
n
Then switch on the converter with the
START/
STOP
button.
Hold the top section of the PSKP syringe (E4)
n
with two fingers for approx. 10 seconds. This
action accelerates the cold plasma conversion
process in the air chamber. The patient probe
has to be maintained during these 10 seconds in
user's hand.
n
After approx. 10 seconds, switch off the converter
at the converter or the energy control unit,
because the filling process is now complete.
n Now remove the PSKP syringe from the con-
verter and inject it into the relevant treatment
area.
Using the PSKP syringe
The versatility of the various instrument probes will
only be used in dental medicine.

Problem solutions

6.1

Solution table

Problem
Überprüfung
The instrument probe is not active, al-
Was the Start/Stop button pressed for
though the treatment unit is switched
at least 3 seconds?
on.
Is the converter connected to the en-
ergy control unit?
Is the instrument probe damaged in any
way? If yes...
Is the patient holding the patient probe
during treatment?
The treatment unit switches off if the
Has the glass of the instrument probe
instrument probe comes into contact
been damaged during the treatment?
with the patient...
Patients can store more current or can
generate a greater electronic charge
than has been preprogrammed as
standard.
This is reported to the energy control
unit by the patient probe.
Did the end of the instrument probe
which is plugged into the converter be-
come wet during the treatment?
The treatment unit switches off auto-
Is there liquid or other material in the
matically after charging of the PSKP...
probe?
Has the probe been properly plugged
into the handpiece?
Lösung
Press the START/STOP button on the
energy control unit or directly on the
converter until the power indicator light
illuminates!
Check whether the converter plug at the
rear of the energy control unit is prop-
erly plugged into the socket! The arrow
on the plug must be visible at the top!
Replace the defective instrument probe!
Please plug the plug connector of the
patient probe into the rear of the energy
unit and place the probe in the patient's
hand.
Remove the broken glass from the pa-
tient and dispose of the defective instru-
ment probe in a proper manner!
Restart treatment by setting the inten-
sity 1 to 2 levels lower and extend the
treatment time!
Dry the entire instrument probe, espe-
cially the metal piece!
Replace the PSKP with a new syringe!
Reinstall the PSKP!
plasma ONE · Instruction Manual