For installing the lamp, the housing cover (Figure 1[1]) of the compact light source must be
removed. For this, remove the two screws (Figure 1[2]) with an Allen wrench WAF 3.0. If a
lamp is fitted in the Compact Light Source HXP 120 V, that lamp must be removed before
a new lamp can be installed. Remove the lamp in reverse order of the installation
described here.
1
Lever
2
Pressure bolt
3
Lamp holder
4
Protective filter
Figure 2: Device with housing cover removed; no lamp installed
For installing the lamp, pull back the pressure bolt (Figure 2[2] with the lever (Figure 2[1])
(move the lever towards the front panel) and install the lamp. Make sure that the groove in
the contact face of the lamp engages the respective nose in the lamp holder (Figure 2[3]).
If the lamp makes full contact with the lamp holder (Figure 2[3]), the lever (Figure 2[1]) can
42 von 52