Herunterladen Inhalt Inhalt Diese Seite drucken

Gima GIMACARE R-C4A Bedienungsanleitung Seite 39

Vorschau ausblenden Andere Handbücher für GIMACARE R-C4A:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 93
RF e m i s s i o ns
CISPR11
RF e m i s s i o ns
CISPR11
Harmonic emis-
sions
IEC 61000-3-2
Voltage fluctua-
tions/
Flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration — electromagnetic
immunity
The device is intended for use in the electromagnetic environment
assure that it is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
Group 1
likely to cause any interference in
nearby electronic equipment.
Class B
The device is suitable for use in all es-
Not appli-
tablishments including those directly
cable
connected to the public low-voltage
power supply network that supplies
to buildings power used for domestic
Not appli-
purposes
cable
IEC 60601
Test level
±8kV direct
& indirect
contact;
±15kV air
discharge
±2 kV for
power
supply lines
± 1 kV line(s)
to line(s)
Compliance
level
±8kV direct
& indirect
contact;
±15kV air
discharge
not applica-
ble
not applica-
ble
37
Electromagnetic
environment-guidance
Floors should be
wood, concrete or
are covered with
synthetic material,
the relative humidity
should be at least 30%
not applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
not applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
Inhaltsverzeichnis
loading

Diese Anleitung auch für:

28405

Inhaltsverzeichnis