13) Do not use the stimulator in the vicinity of shortwave or
microwave therapy equipment, since this may affect
the output power of the stimulator.
14) Never use it near the cardiac area. Stimula-
tion electrodes should never be placed any-
where on the front of the thorax (marked by
ribs and breastbone), but above all not on
the two large pectoral muscles. There it can increase
the risk of ventricular fibrillation and lead to cardiac ar-
rest.
15) Never use it on the eye, head and face area.
16) Never use it near the genitals.
17) Never use it on the areas of the skin which lack normal
sensation
18) Keep electrodes separate during treatment. It could re-
sult in improper stimulation or skin burns if electrodes
are in contact with each other.
19) Keep the stimulator out of reach of children.
20) Consult your doctor if you are in any doubt whatsoever.
21) Discontinue it and do not increase the intensity level if
you feel discomfort during use.
2.2.3
Precautions
1) TENS is not effective for pain of central origin including
headache.
TENS is not a substitute for pain medications and other
2)
pain management therapies.
TENS is a symptomatic treatment and, as such, sup-
3)
presses the sensation of pain that would otherwise
serve as a protective mechanism.
Effectiveness is highly dependent upon patient selec-
4)
tion by a practitioner qualified in the management of
11