④ If still fails to start, contact your dealer or local service center.
●No waveform displayed after signal acquisition:
① Check probe connection to signal cable.
② Check signal cable connection to BNC (channel connector).
③ Check probe connection to device under test.
④ Confirm device under test is generating signal.
⑤ Re-acquire the signal.
●Measured voltage amplitude differs from actual (common with probe use):
① Verify channel attenuation matches probe attenuation ratio.
●Waveform appears stepped:
① Horizontal time base may be too low; increase to improve resolution.
② If display type is Vector (lines between samples), switch to Point mode to remove stepped effect.
5.MAINTENANCE
5.1 Maintenance
When storing or placing the instrument, avoid exposing the LCD screen to direct sunlight for extended
periods.