Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The BC30 is intended for use in the electromagnetic environment specified below. The customer or the user of BC30 should assure that it is used
in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should be used no closer to any part of
the BC30 , including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 3.5
3
Vrms
V
d = 3.5
80 MHz to 800 MHz
E
1
3
Vrms
d = 7
E
1
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range
may occur in the vicinity of equipment marked with the following symbol:
80 MHz to 2.5 GHz
93
a
.b
. Interference