USE OF NEBULIZATION MANUAL CONTROL
To achieve continuous nebulisation action you should not use the manual nebulisation control (B6),
especially in the case of children or persons with reduced physical, sensory, or mental capabilities. The
manual nebulisation control is useful for limiting dispersion of the medication in the surrounding en-
vironment.
SoftTouch masks
otect
lution
-
cult.
er part
HYGIENIC PREPARATION
Switch o the appliance before each hygienic preparation operation and disconnect it from the mains.
Compressor unit (A) and pipe exterior (B1)
Use only a cloth moistened with antibacterial detergent (non-abrasive and free of solvents of any kind).
Nebuliser and accessories
To disassemble the bowl, rmly grasp the bowl body and then use your other hand to rotate the particle
size regulator (B2.4) clockwise until it is ejected from the upper part (B2.3) as a result of unscrewing.
The components can then be disassembled (regulator B2.4a, valve B2.4b, selector B2.4c and upper part
B2.3) to be cleaned. The nozzle (B2.2) is interference tted in the upper part (B2.3) and can be extracted
by simply pulling it outwards.
Then proceed according to the following instructions.
Sanitisation
Before and after each use, sanitise the nebuliser and accessories by choosing one of the methods
provided in the table and described below.
method A: Sanitise the accessories under warm drinking water (approx. 40°C) with mild dishwashing
detergent (non-abrasive).
method B: Sanitise the accessories in the dishwasher with the hot cycle (70°C).
method C: Sanitise the accessories by soaking in a solution of 50% water and 50% white vinegar, then
rinse thoroughly with warm drinking water (approx. 40°C).
After sanitising the accessories, shake them vigorously and place them on a paper towel, or
alternatively dry them with a hot air jet (e.g. hair dryer).
Disinfection
After sanitising the nebuliser and accessories, disinfect them using one of the methods provided in the
table and described below. Each method can be performed for a limited number of times (see gure
in table).
method A: Obtain a disinfectant of the electrolytic chloride type (active ingredient: sodium
To start nebulizing close with a
nger the hole of the nebulizer
manual control (B6) and breathe
in gently; we recommend to
hold your breath for a moment
so that the inhaled aerosol
droplets can be deposited,
SoftTouch masks have an outer edge made of soft, biocompatible ma-
Soft
terial that ensures an optimal t on the face, and are also equipped with
biocompatible
theinnovative Dispersion Limiter. These characteristic features allow
material
for greater sedimentation of the drug in the patient, and again limit its
dispersion.
Dispersion
Limiter
Upon inhalation, the tab, which acts as
a Dispersion Limiter, bends inwards in
the mask.
meanwhile, to disable nebuliz-
ing, remove your nger from the
hole of the nebulizer manual
control to avoid the waste of
drug, optimizing its acquisition.
Then exhale slowly.
Upon exhalation, the tab, which acts as
a Dispersion Limiter, bends outwards
from the mask.
13