Program 23 + u2 :
Program 24 + u24:
Program 25 + u25:
schwa-medico OPErATING MANuAL - EMP 4 ECO+
Combination: Active knee joint motion + pain management
Type of stimulation
Muscle stimulation and gate control
Parameter
Frequency channel
Frequency channel
Pulse width: 300 μs
Rising ramp: 5 s
Working phase: 1 s
Falling ramp: 5 s
Pause: 10 s
Time: 20 min
Description
See picture program P21
Combination: Active ankle joint motion + pain management
Type of stimulation
Muscle stimulation and gate control
Parameter
Frequency channel
Frequency channel
Pulse width: 300 μs
Rising ramp: 3 s
Working phase: 1 s
Falling ramp: 1 s
Pause: 6 s
Time: 20 min
Description
See picture program P21
Combination: hemiparesis (shoulder) + pain management
Type of stimulation
Muscle stimulation and gate control
Parameter
Frequency channel
Frequency channel
Pulse width: 250 μs
Rising ramp: 2 s
Working phase: 6 s
Falling ramp: 2 s
Pause: 10 s
Time: 20 min
Description
See picture program P21
(black) and
(red): 45 Hz
(grey) and
(yellow): 90 Hz
(black) and
(red): 45 Hz
(grey) and
(yellow): 90 Hz
(black) and
(red): 45 Hz
(grey) and
(yellow): 90 Hz
57