Herunterladen Inhalt Inhalt Diese Seite drucken

schwa-medico 101707 Gebrauchsanweisung Seite 55

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Program 18 + u18:
Program 19 + u19:
intensity
T = 1/f
pulse width
Program 20 + u20:
channel 2
intensity
Channel 1
frequency
Hz
100
phase 1
50
2
10
frequency
Hz
100
phase 1
50
schwa-medico OPErATING MANuAL - EMP 4 ECO+
Mixed Incontinence
Type of stimulation
Parameter
Description
Stress Incontinence
Type of stimulation
Parameter
Description
relaxation – Dynamic Stimulation
Type of stimulation
time
Parameter
Symbols on the
display during
stimulation
time
Description
phase 2
time
min
20
30
Muscle stimulation
Frequency: 20 Hz
Pulse width: 180 μs
Rising ramp: 2 s
Working phase: 4 s
Pause: 4 s
Time: 15 min
See picture program P12
Muscle stimulation
Frequency: 50 Hz
Pulse width: 180 μs
Rising ramp: 2 s
Working phase: 3 s
Pause: 6 s
Time: 15 min
See picture program P12
frequency
Hz
100
build-up
phase
run-down phase
50
Dynamic stimulation (slow)
Frequency: 120 Hz
7,5
Pulse width: 150 μs
Time: 20 min
intensity
mA
Rising ramp: 2 s
70
working phase
Falling ramp: 2 s
Rising phase
Falling phase
The 4 channels are working with the same frequency
2
and pulse width.
intensity
intensity I
mA
T = 1/f
200 µs
75 µs
run-down
phase
time
s
15
pause
phase
time
s
7
8
time
55

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Emp 4 eco+ 101060

Inhaltsverzeichnis