English
|
Assembly
4
Assembly
4.1
Mounting on the wall without the angled plate
surface as described here.
Fig. 5
NOTICE
Damage to the device
If the carrying system holder is not fixed tightly enough, it may loosen and fall.
► Always mount the carrying system holder with the 4 screws provided.
► The assembly must be performed by specialized personnel only.
mounted, to verify sufficient sturdiness. For further information, see chapter "Technical data".
1. Make sure that the surface is sturdy enough.
2. Drill the 4 fixation holes according to the dimension drawing.
3. Screw the base plate to the wall with 4 M6 screws.
4. Insert the tie rod into the slot on the release lever and slide it all the way to the left.
5. Place the cover on the base plate and screw it down with the 4 M5 screws provided.
6. Optional: Connect an external power supply source to the DC power supply connector.
20
Instead of a wall, the carrying system holder can also be mounted on a flat, horizontal
140
Dimension drawing for wall mounting (measurements in mm)
Before mounting, check the surface, onto which the carrying system holder is to be
Only mount the carrying system holder horizontally with the release lever on the left side.
Instructions for use
125
125
|
Carrying system holder for Oxylog VE300