IEC 60601-1-2:
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge(ESD) IEC
60601-4-2
Electrical fast transient burst IEC
61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4- 11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
IEC 60601-1-2:
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should ensure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-67
Radiated RF IEC 61000-4-3
IEC 60601 test level
± 8 kV Contact
± 15 kV Air
± 2 kV power cord
± 1 kV input/output
± 0.5 kV and ± 1kV line(s) to
line(s).
0% 230 V for 1 cycle
0% 230 V for 0.5 cycle
at 8 phase angles
70% 230 V (30% dip in 230V) for
25/30 cycles
0% 230V for 250/300 cycles
30 A/m
IEC 60601 test level
6 Vrms 150 kHz - 80 MHz
10 V/m 80MHz-2,7GHz
Compliance level
± 8 kV Contact
± 15 kV Air
± 2 kV power cord
± 1 kV input/output
± 0.5 kV and ± 1 kV line(s) to
line(s).
0% 230 V for 1 cycle
0% 230 V for 0.5 cycle
at 8 phase angles
70% 230 V (30% dip in 230V) for
25/30 cycles
0% 230V for 250/300 cycles
30 A/m
Compliance level
6 Vrms 150 kHz - 80 MHz
10 V/m 80MHz-2,7GHz
EN