8.3.5 :SENSe:SWEep:GENeration <STEPped | ANALog>;
:SENSe:SWEep:GENeration ?;
Description: Sets or returns the signal display regarding the signal recording mode.
Forms:
Parameters: STEPped, ANALog.
*RST Value: N/A.
8.3.6 :SENSe:VOLTage] <f | MAXimum | MINimum> [V];
:SENSe:VOLTage] <f | MAXimum | MINimum> [V];
:SENSe:VOLTage] [MAXimum | MINimum] ?;
:SENSe:VOLTage] [MAXimum | MINimum] ?;
Description: Sets or queries the peak-to-peak input voltage range (in volts). To set or
Parameter: A floating point number, which can optionally be completed by a unit (V, mV,
Forms:
*RST Value: N/A.
Examples:
8.4 SELect Subsystem
8.4.1 :SELect[:CATalog] ?;
Description: Returns a comma separated character string, containing all trace names
Forms:
Parameters: N/A
*RST Value:N/A
8.4.2 :SELect:<trace_name> <ON I OFF>;
:SELect:<trace_name> ?;
Description: This command switches a channel regarding signal display on or off.
Forms:
Parameters: N/A.
*RST Value: N/A.
Example:
8.5 Trigger Subsystem
8.5.1 :INITiate[:IMMediate];
Description: This command resets the scope if it is in digital and single mode
5 0
"STEPped" stands for digital and "ANALog" for analog mode.
Command and query.
query the peak-to-peak input voltage range for the second channel, append
a suffix of 2 to the VOLTage header. This range is displayed in 10 divisions.
µV). MAXimum and MINimum enable to set or return the minimum or
maximum input deflection coefficient.
Command and query.
":SENSE:VOLTAGE 100mv;" // sets input CH I to 10mV/div.
":sense:volt minimum;" // sets CH II to the minimum deflection coefficient.
":SEN:volt2 ?;" // query for CH II deflection coefficient. If the minimum
deflection coefficient is present, the response is "1.00E-2;" 10mV for 10
division (1mV/div).
(trace name).
Query.
Queries show the trace (channel, reference memory) status.
Query.
":sel:ch2 off;" // switches channel II off.
(SENSe:SWEep:GENeration STEPped, TRIGger:DELay:AUTO ONCE). This
means the trigger system is armed.
Änderungen vorbehalten / Subject to change without notice