S S p p e e c c i i f f i i c c c c o o n n d d i i t t i i o o n n s s o o f f u u s s e e
A A . .
P P M M A A . . E E v v o o l l u u t t i i o o n n a a n n d d Y Y C C O O 1 1 4 4 - - Y Y E E x x - - L L i i n n k k C C o o n n v v e e r r t t e e r r
1)
The enclosures of the PMA.Evolution and of the YCO14-Y Ex-Link Converter are non-
conducting and may generate an ignition-capable level of electrostatic charges under
certain extreme conditions. The user shall ensure that the equipment is not installed in a
location it where may be subjected to external conditions that might cause a build-up of
electrostatic charges on non-conducting surfaces, additionally; cleaning of the
equipment should be done only with a damp cloth.
5)
The PMA.Evolution shall only be used with the YCO14-Y Ex-Link Converter and vice
versa. The CAT6 cable used between the PMA.Evolution and the YCO14-Y Ex-Link
Converter shall be obtained from Sartorius: Cable type YCC01-0052Mxx (XX = 10 for
10m, 20 for 20m; 30 for 30m); maximum cable length: 100ft (30.5m). The data cable
connected to the scale (weighing unit) is considered as intrinsically safe circuit. Check
the correct function of the data transfer before you use the equipment in a hazardous
location. The equipment operator is responsible for any non-Sartorius cables used.
B B . .
Y Y C C O O 1 1 4 4 - - Y Y E E x x - - L L i i n n k k C C o o n n v v e e r r t t e e r r
4)
The protective ground connectors on the PMA.Evolution and the YCO14-Y Ex-Link
Converter shall be connected together using a potential equalisation conductor (PE)
with a cross-sectional area of ≥ 4mm². The low resistance of this connection to the PE bus
bar must be checked when the system is installed at the intended place of use. The
shielding of the connecting cables may only be used for grounding when no
impermissible difference in voltage is generated and, if necessary, the shielding is able to
conduct the equipotential current.
6)
If the Ex-Link Converter YCO14-Y is installed in Zone 2, the installation must be done in
such a way that the pollution degree 2 of the YCO04-Y board is guaranteed.
8)
In Zone 2 the „USB Power" port may not be used.
9)
If the cable contains no Ex ic circuits, the cable must be protected against unintended
release by use of Sartorius „Strain Relief for USB Cord" type Y Y S S R R 0 0 1 1 .
10) The equipment shall be installed in such a way that it is protected against UV radiation
and the entry of solid foreign objects or water capable of impairing the safety of the
apparatus. Reduce the risk of mechanical damage to a minimum.
2021-04-28
Dr. D. Klausgrete
58
EVO1X | LAB1X Installation Instructions