5 Hygiene treatment
5 Hygiene treatment
5.1 General information
5.2 Intervals
5.3 Subjecting module to a hygiene
Requirement
Part
Housing
Network cable
44
EN
prismaCONNECT
•
Wear appropriate safety gear for disinfection.
•
Follow the instructions for use of the disinfectant used.
Interval
Weekly
If required, e.g., following infectious
diseases or severe contamination
treatment
Material damage due to ingress of liquids!
Ingress of liquids may damage therapy device,
components, and accessories.
Do not immerse therapy device, components, and
accessories in liquids.
The therapy device is switched off (see instructions for use of
the therapy device).
1. Disconnect the module from the therapy device
Removing the prismaCONNECT module", page
2. Subject the module to a hygiene treatment in accordance
with the table below:
Cleaning
Wipe down:
Use water or mild
detergent
Disinfection
Disinfect by wiping
(Recommended:
®
terralin
protect or
perform advanced
Alcohol EP)
Action
Clean modules
Disinfect modules
(see "4.2
41).
Sterilization
Not permitted