Error codes (HS / HP)
Any malfunctions during operation will be identified by an error message on the display (C).
Proceed as follows in such cases:
- Whenever an error message appears switch the device off to cool down
- Carry out corrective measures
- Restart device
Error code
Cause
Solution
E1
Temperature inside device too high
- Switch off device and allow to cool down
E6
Motor blockage
- Stirring bars inappropriate
- Reduce the viscosity
- Remove foreign objects in the device
E9
Safety circuit ETC
- Plug the contact plug
- Plug the contact thermometer
- Replace the defective connecting cable, plug or contact thermometer
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the fault.
Maintenance and cleaning
The device is maintenance-free.
/// Cleaning
- For cleaning disconnect the mains plug!
- Use only cleaning agents which have been approved by IKA to clean the devices:
Water containing surfactant / isopropyl alcohol.
- Wear protective gloves during cleaning the devices.
- Electrical devices may not be placed in the cleansing agent for the purpose of cleaning.
- Do not allow moisture to get into the device when cleaning.
- Before using another than the recommended method for cleaning or decontamination, the user must ascertain with IKA that this
method does not destroy the device.
16
/// Spare parts order
When ordering spare parts, please give:
- device type.
- serial number, see type plate.
- position number and description of spare part, see www.ika.com.
/// Repair
Please only send devices in for repair that have been cleaned and are free of materials which might present health hazards.
For repair, please request the "Safety Declaration (Decontamination Certificate)" from IKA or use the download printout of it from
IKA website at www.ika.com.
If your appliance requires repair, return it in its original packaging. Storage packaging is not sufficient when sending the device - also use
appropriate transport packaging.
Accessories
• For further accessories see www.ika.com.
Assembling the stand
• Remove screw plugs (F).
• Remove the protective cap from the support rod.
• Put the washer between housing and nut.
• Screw the support rod onto the device by hand until the end
stop is reached.
• Use an A/f 17 spanner to tighten the M10 nut.
• Accessories should be attached using cross sleeves.