Error codes
• The fault is shown by an error code on the display as following if the error occurs. (Fig. 5)
• Proceed as follows in such cases:
- Disconnect the power supply for at least 10 sec.
- Carry out corrective measures.
- Restart the device.
Error code Cause
Effect
Er 1
Sensor not connected or defective (Sensor
Device off
rupture)
Er 5
Sensor not in media (no change in tempe-
Device off
rature of the media after heater active for
3 minutes)
Er 6
Temperature sensor short-circuit
Device off
Er 7
The temperature of the medium is higher
Device off
than the maximum target temperature that
can be set
Er 8
Calibration error
Device off
• If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the fault.
Accessories
• For accessories see www.ika.com.
16
Solution
- Connect temperature sensor
- Replace temperature sensor
- Insert the temperature sensor in the media to a depth of at
least 20 mm
- Switch off the device and replace temperature sensor
- Switch off the device and allow the media to cool down
- Increase the maximum target temperature that can be set
- Switch off the device
- Re-calibration by authorised service personnel
Technical Data
Temperature measuring range
°C
Temperature measurement resolution
K
Accuracy of temperature measurement
K
Set temperature resolution
K
Type of control
Control deviation (500 ml water in 600 ml glass beaker, mag-
K
netic stirring bar 40, at 600 rpm, 50 °C)
Immersion depth max.
mm
Supply voltage
V DC
Current consumption
mA
Permissible ambient temperature
°C
Permissible relative humidity
%
Permissible duration of operation
%
Protection class according to DIN EN 60529
Cable length
mm
Male connector
Socket
Dimensions (W x D x H)
mm
Weight
kg
Subject to technical changes!
Warranty
In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim. You
will be liable for freight costs.
- 50 ... 450
0.1
± 0.2 + sensor tolerance PT 1000
DIN IEC 751 Class A
0.1
PID
± 0.5
200
8 ... 16 (Limited Power Supply)
15
0 ... 60
80
100
IP 54
350
6-pin DIN 45322
DIN 12878 Class 2
82 x 22 x 83 (without sensor)
0.2
The warranty does not cover worn out parts, nor does it apply
to faults resulting from improper use, insufficient care or mainte-
nance not carried out in accordance with the instructions in this
operating manual.
17

Werbung

loading