ARGON MEDICAL DEVICES, INC.
1445 Flat Creek Road
Athens, Texas 75751 USA
Tel: 800-927-4669
Tel: +1 903-675-9321
www.argonmedical.com
Emergo Europe, Molenstraat 15, 2513 BH.
The Hague, The Netherlands
Tel: +31 (0)70 345 8570
Fax: +31 (0)70 346 7299
0086
IFU7503/0114
SKATER™ Nephrostomy Kit Non-Locking
1 / 18
en - English
Indication for use
The product is for drainage using Seldinger technique
Duration
Up to 12 weeks
Warnings
This device was designed, tested and manufactured
for single use only. Reuse or reprocessing has not
been evaluated and may lead to its failure and
subsequent patient illness, infection or other
injury. Do not reuse, reprocess or re-sterilize this
device.
Do not use if package appear open or if the expiry
date has been exceeded.
The product must only be used by qualified
personnel who are familiar with the technique.
Preparation
Loosen the straightener together with the guidewire.
Advance the straightener carefully over the
curvature of the catheter, while straightening the
curvature with your fingers.
Introduce the chosen stiffener (if plastic: activate the
hydrophilic coating with saline) and fasten the
stiffener onto the catheter (Luer Lock).
Remove the straightener
Activate the coating by saline.
Procedure
Perform the skin incision under local anaesthesia.
Introduce the needle into the cavity in question using
ultrasound guidance, CT or fluoroscopy.
Verify correct position by aspiration or by injection of
contrast medium and fluoroscopy.
When correctly positioned introduce the guidewire
through the needle.
Withdraw the needle over the guidewire.
Dilate the canal.
Advance the catheter over the guidewire in
fluoroscopy.
When the catheter tip is inside the cavity, loosen the
stiffener.
Hold the stiffener tight and advance the catheter
over the stiffener until the complete length of the
pigtail is in the cavity.
Remove the stiffener.
Verify correct position of the catheter using
fluoroscopy.
Remove the guidewire
Connect the catheter to a drainage bag using a
suitable connecting tube.
It is recommended affixing the catheter in a straight line
and that any curvature be applied to the connecting tube.
It is recommended that the physician follow the
hospital's standard of care procedures for drainage
catheters.