Herunterladen Inhalt Inhalt Diese Seite drucken

Dmax; Dmin; Ddiff; Dmean - marposs TESTAR Quick read M Installations- Und Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 65
Example :
Request of the Nominal Value on QuickRead with selected metric system, low
resolution and NominalValue = 000.008
Request from Host
Response from QuickRead

DMAX

Request from Host
Response from QuickRead
Select Dynamic Type Measure of Maximum.
The QuickRead checks and selects dynamic type measure of maximum and returns ↵ .
In case of error QuickRead returns the string ERR3
The QuickRead display a NAN value how digital number and the bargraph is off.

DMIN

Request from Host
Response from QuickRead
Select Dynamic Type Measure of Minimum.
The QuickRead checks and selects dynamic type measure of minimum and returns ↵ .
In case of error QuickRead returns the string ERR3
The QuickRead display a NAN value how digital number and the bargraph is off.

DDIFF

Request from Host
Response from QuickRead
Select Dynamic Type Measure of Maximum - Minimum.
The QuickRead checks and selects dynamic type measure of maximum - minimum and
returns ↵ .
In case of error QuickRead returns the string ERR3
The QuickRead display a NAN value how digital number and the bargraph is off.

DMEAN

Request from Host
Response from QuickRead
Select Dynamic Type Measure of (Maximum + Minimum)/2.
The QuickRead checks and selects dynamic type measure of (maximum – minimum)/2 and
returns ↵ .
In case of error QuickRead returns the string ERR3
The QuickRead display a NAN value how digital number and the bargraph is off.

DSEMI

Request from Host
Response from QuickRead
198
NOMVAL?
000.008
DMAX
↵ .
DMIN
↵ .
DDIFF
↵ .
DMEAN
↵ .
DSEMI