Technical Information
RS 232/RS 485 Communication Description
When a data is indicated between two quotes ('...') it means that the indicated
NOTE
data is the corresponding ASCII character.
Data Type
Logic (L)
Numeric (N)
Alphanumeric (A)
260/276
<STX> (Start of transmission) = 0x02
<ADDR> (Unit address) = 0x80 (for RS 232)
<ADDR> (Unit address) = 0x80 + device number (0 to 31) (for RS
485)
<WIN> (Window) = a string of 3 numeric char-acter indicating the
window number (from '000' to '999'); for the meaning of each
window see the relevant paragraph.
<COM> (Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA> = an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present.
<ETX> (End of transmission) = 0x03
<CRC> = XOR of all characters subsequent to
<STX> and including the <ETX> terminator. The value is
hexadecimal coded and indicated by two ASCII character.
Field Length
1
6
10
TV 902 Navigator User Manual / 87-900-999-01 (B)
Valid
Characters
'0' = OFF
'1' = ON
'-', '.', '0' . . . '9' right
justified with '0'
from blank to '_' (ASCII)