Herunterladen Diese Seite drucken

Agilent Technologies 8698933R001 Bedienungsanleitung Seite 250

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
17
Technical Information
Interconnections
PROGRAMMABLE ANALOG SIGNAL: this output signal is a voltage
(from 0 to 10 Vdc) proportional to a reference quantity (frequency or
power) set by the user. The default setting is the frequency (see the
following example diagram).
Figure 11
FAULT: this open collector output signal is ON when a system fault
condition is detected.
PROGRAMMABLE SET POINT: this open collector output signal is
enabled when the reference quantity chosen (frequency, current or
time) is higher than the set threshold. The signal can be "high level
active" (that is the output is normally at 0 Vdc and becomes 24 Vdc
when activated), or "low level active" (that is the output is normally
at 24 Vdc and becomes 0 Vdc when activated). Moreover, if the
reference quantity is the frequency or the current drawn, it is
possible to set the hysteresis (in % of the threshold value) to avoid
bouncing.
For example:
250/276
reference quantity: frequency
threshold: 500 Hz
hysteresis: 1%
activation type: "high level"
TV 902 Navigator User Manual / 87-900-999-01 (B)

Werbung

loading

Diese Anleitung auch für:

Tv 902 navigator