Measuring principle
Temperature compensation
Measuring conditions
Meas. range
Temperature range
Max. admissible overpressure
Immersion depth
Operating position
Incident flow
Connections
Characteristics when delivered
Zero signal
Response time at 20°C
Self consumption at 20°C
Drift
Operation time
Material
Dimensions
70
Technical data
Membrane covered galvanic probe
IMT
0 - 50 mg/l O 2
0°C - 50°C
6 bar
min.
6 cm
max. 20 m depth
any
> 3 cm/s at
10 % meas. accuracy
10 cm/s at
5 % meas. accuracy
18 cm/s at
1 % meas. accuracy
Cable, plug:
Protection degree IP67, 8pin plug
Probe:
Protection degree IP68 (20 m)
< 0.1 % of saturation value
t 90 (90 % of indication of end value after) < 10 s
t 95 (95 % of indication of end value after) < 16 s
t 99 (99 % of indication of end value after) < 60 s
0.008 µg h -1 (mg/l) -1
approx. 3 %/month in operation condition
at least 6 months per electrolyte filling
Membrane head
Membrane
Probe head
Thermistor housing
Shaft
145 mm
15.25 mm
13 µm
1.5 m to 20 m
POM
FEP
POM
VA-steel 1.4571
POM
shaft length
shaft diameter
membrane thickness
cable length