Herunterladen Diese Seite drucken

OHAUS Frontier FC5720R Bedienungsanleitung Seite 35

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-28
9.1.4 Centrifuge FC5720R
Model
Speed Range
Maximum RCF
Maximum Capacity (Rotor)
Temperature range (Digital)
Running Time
Noise level (depending on the rotor)
Allowable density at maximum speed
Allowable kinetic energy
Mains power connection AC
Voltage fluctuation
Current consumption
Power consumption
Dimensions (W × D × H)
Net Weight (without rotor)
Shipping Dimensions (W × D × H)
Shipping Weight (without rotor)
Ambient conditions (EN/IEC 61010-1)
Environment
Altitude
Ambient temperature
Max. relative humidity
Overvoltage category (IEC 60364-4-443)
Degree of contamination
Class of protection
Not suitable for use in hazardous environments.
FC5720R
200 rpm -20000 rpm;10 rpm/set
38007 x g;10 x g/set
4 x 200 ml
-20° to 40°C/1°C increments
10 sec to 99 hr 59 min 59 sec or continuous
≤ 60 dB(A)
1.2 g/ml
24367 Nm
230 V ~ 50/60 Hz
± 10 %
5.9 A
1.2 kW
407 x 712 x 361 mm
16.0 x 28.0 x 14.2 in
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
for indoor use only
Use up to an altitude of 2000 m
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
TM
Frontier
Centrifuges
120 V ~ 50/60 Hz
± 10 %
10.5 A
1.1 kW
61 kg
157 lb
83 kg
183 lb
II
2
I

Werbung

loading