Centrifuges
9.1.3 Centrifuge FC5718R
Model
Speed Range
Maximum RCF
Maximum Capacity (Rotor)
Temperature range (Digital)
Running Time
Noise level (depending on the rotor)
Allowable density at maximum speed
Allowable kinetic energy
Mains power connection AC
Voltage fluctuation
Current consumption
Power consumption
Dimensions (W × D × H)
Net Weight (without rotor)
Shipping Dimensions (W × D × H)
Shipping Weight (without rotor)
Ambient conditions (EN/IEC 61010-1)
Environment
Altitude
Ambient temperature
Max. relative humidity
Overvoltage category (IEC 60364-4-443)
Degree of contamination
Class of protection
Not suitable for use in hazardous environments.
FC5718R
200 rpm -18000 rpm;10 rpm/set
23542 x g;10 x g/set
4 x 200 ml
-20° to 40°C/1°C increments
10 sec to 99 hr 59 min 59 sec or continuous
≤ 60 ± 2 dB(A)
1.2 g/ml
25111 Nm
230 V ~ 50/60 Hz
± 10 %
3.0 A
660 W
407 x 731 x 359 mm
16.0 x 28.8 x 14.1 in
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
for indoor use only
Use up to an altitude of 2000 m
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
120 V ~ 50/60 Hz
± 10 %
6.0 A
660 W
60 kg
132 lb
77 kg
170 lb
II
2
I
EN-27

Werbung