Centrifuges
9. TECHNICAL DATA
9.1 Specifications
9.1.1 Centrifuge FC5714
Model
Speed Range
Maximum RCF
Maximum Capacity (Rotor)
Temperature range (N/A)
Running Time
Noise level (depending on the rotor)
Allowable density at maximum speed
Allowable kinetic energy
Mains power connection AC
Voltage fluctuation
Current consumption
Power consumption
Dimensions (W × D × H)
Net Weight (without rotor)
Shipping Dimensions (W × D × H)
Shipping Weight (without rotor)
Ambient conditions (EN/IEC 61010-1)
Environment
Altitude
Ambient temperature
Max. relative humidity
Overvoltage category (IEC 60364-4-443)
Degree of contamination
Class of protection
Not suitable for use in hazardous environments.
FC5714
200 rpm -14000 rpm;10 rpm/set
18624 x g;10 x g/set
4 x 200 ml
10 sec to 99 hr 59 min 59 sec or continuous
≤ 63 ± 2 dB(A)
1.2 g/ml
5595 Nm
230 V ~ 50/60 Hz
± 10 %
1.3 A
240 W
362 x 493 x 330 mm
14.3 x 19.4 x 13.0 in
580 x 490 x 460 mm
22.8 x 19.3 x 18.1 in
32.5 kg
For indoor use only
Use up to an altitude of 2000 m
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Air cool
120 V ~ 50/60 Hz
± 10 %
2.4 A
300 W
30 kg
66 lb
72 lb
II
2
I
EN-25