Herunterladen Diese Seite drucken

Endress+Hauser Deltabar S PMD70 Austauschanleitung Seite 7

Austausch der sensorbaugruppe
Vorschau ausblenden Andere Handbücher für Deltabar S PMD70:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Exchange of sensor unit on Deltabar S PMD70
0
2
1
Before de-installation, it has to be made sure that the supply voltage for the device is switched off
housing T14, T15, T17
as example: housing T14
1
T14/T15:
6
retaining ring
T17:
locking ring
7
sensor unit
PMD70
Dismantle instrument, disassemble sensor unit
1
Unscrew lid from terminal compartment, disconnect cable from terminal, uninstall instrument.
Unscrew lid from electronic compartment.
2
3
If installed remove display from electronic module:
• software version 1.x
remove display from holder by pressing the hooks outwards, unscrew both screws and remove holder.
³
software version
2.x
remove display from electronics-module by pressing the hooks outwards.
4
Unplug the display cable from electronic module.
5
Loosen mounting screws at electronic module and pull module out of the housing, plug connection!
Separation of housing and sensor unit
6
Loosen set screw about 1-2 turns.
7
Clamp housing in a vice, take care not to damage the housing. Screw off the sensor unit counterclockwise from housing (9 turns);
it is not easy to unscrew the housing because the retaining ring is blocking - it has to be exchanged afterwards.
Attention: Take care not to damage the sensor cable!
KA603P/00/a2/12.06
5202 1845
Ausgabe/revision: 3
The instrument may only be maintained and repaired by qualified personnel.
and legal requirements as well as any certificates have to be observed!
Only modular assemblies may be exchanged against identical, original Endress+Hauser spare parts !
5
electronic module
software 1.x
display cable
sensor
4
cable
holder
electronic module
³
software
2.x
3
4
sensor cable
3
display
lid
electronics compartment
The device documentation, applicable standards
The exchange requires the following tools:
• Vice
• Philips screw driver size 1
• Allen wrench AF2
• open ended or ring spanner AF17
• Torque wrench 30 Nm
display cable
2

Werbung

loading