Individual Settings
Various factory settings on the measuring amplifier card can be altered. Adap-
tations, which are of no problem, can be carried out by means of the DIP
switch on the board or also by externally wired switching elements. If re-
quested with the order, they can be carried out by the manufacturer.
4.1 Operational Setting
S21
Switch
The transducer signal is switched off in the "Zero" function, so that the "Zero
Value" of the amplifier is applied to the output. This function enables amplifiers
to be replaced in measuring chain without the need for rebalancing the bridge
with the unloaded transducer. All that is required is that the amplifier balance
values are noted and then this value is set, if necessary, with P23, S22 and
S23.
4.2 Setting the Measuring Range
Measuring range 1 is get factory setting. Measuring range 2 can be activated
by switching DIP switch S24/4 or by external switching.
Measuring Range
MR 1 = "2 mV/V
(factory settings)
MR 2 = "0.2 mV/V
Any alteration to the measuring ranges between "0.2 mV/V and "4 mV/V
with U
= 5 V (resp. "0.4mV/V and "8 mV/V with U
B
by means of the DIP switches S26 and S27. For fine balancing see 3.3.
HBM
Zero
Measure
or connect pin 21a to pin 15c
or connect pin 25a to pin 15c
S 24/4
Calibrate
Externally
from MR1 to MR2;
connect 25c and 15c
= 2.5V) can be made
B
A0261−3.1 en/de/fr
ME30