Rohde & Schwarz RT-ZD40 Bedienungsanleitung Seite 69

Werbung

®
R&S
RT-ZD40
As soon as the probe is connected to the oscilloscope and the settings are adjus-
ted, the waveform is shown for the channel to which the probe is connected.
To display probe-specific information and parameters:
1. On the "Vertical" menu, select "Probe Setup".
2. On the left side of the "Setup" tab, select the channel to which the probe is
connected.
The complete probe information is shown on the "Probe Attributes" tab.
2.1.3
AutoZero
This procedure performs an automatic correction of the zero error displayed on
the oscilloscope's screen. The waveform is set to 0 V on the horizontal centerline
of the oscilloscope.
To perform the AutoZero function using the micro button (see also:
"Micro
Button",
on page 18):
1. Connect the probe to the R&S RTO oscilloscope.
2. Configure the micro button function on the R&S RTO:
"Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action" = 'Auto-
Zero'.
3. Short the signal pins and the ground pin together.
4. Press the micro button.
User Manual 1410.5363.02 ─ 03
Putting into Operation
Installation
chapter 2.2,
17