PX Series Balance
5.3.2 Filter Level
Set the amount of signal filtering.
Low
= faster stabilization time with less stability.
Medium
= normal stabilization time with normal stability.
High
= slower stabilization time with more stability.
5.3.3 AZT (Auto Zero Tracking)
Set the automatic zero tracking functionality.
Off = disabled.
0.5 d = display maintains zero up to a drift of 0.5 graduation per second.
1 d = display maintains zero up to a drift of 1 graduation per second.
3 d = display maintains zero up to a drift of 3 graduations per second.
5.3.4 Auto Tare
Set the automatic tare.
Off = disabled.
On = enabled.
Note: "Place container" will be displayed when Automatic Tare is set to On.
5.3.5 Graduations
Set the displayed readability of the balance.
1 Division = standard readability.
10 Divisions = readability is increased by a factor of 10.
For example, if the standard readability is 0.01g, selecting 10 Divisions will result in a displayed reading of 0.1 g.
5.3.6 Date Format
Set the current date format.
YYYY/MM/DD
MM/DD/YYYY
DD/MM/YYYY
5.3.7 Date Setup
Set the date in the current date format.
For example, if the date format is MM/DD/YYYY, the date could be set as "06/22/2017 Thu".
5.3.8 Time Format
Set the current time format.
24HR
12HR
5.3.9 Time Setup
Set the time in the current time format.
For example, if the time format is 24HR, the time could be set as 08:00:00.