Enter the density of the sinker.
4.
Press
and the last digit flashes.
5. Use
to increase the numeric value of the flashing cipher.
Use
to move the number selection to the right, the respective active
position flashes.
6.
Confirm input by
+
Density determination liquids
7.
Place sinker in the upper sample dish.
8.
Press
and the weight of the sinker in air is displayed.
9.
Lay the sinker in the lower sample dish and immerse it into the test liquid.
Make sure that the sinker is at least 1 cm submerged and that, if possible, no
air bubbles adhere to it.
10.
Press
and for a short time the weight of the sinker in the test liquid is
displayed.
The scale determines the density of the liquid and then displays the result.