Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo INGOLD EasyClean 100 Bedienungsanleitung Seite 23

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EasyClean 100/150
h
h
h
© 06 / 2023 METTLER TOLEDO
Printed in the United States
Electrical connections EasyClean 150 with METTLER TOLEDO transmitters
Electrical connections
Transmitter X100 e
NOTE
Bridge terminal 2 and 3 in the EasyClean 150.
Define rinse interval and rinse time according to the
process requirements on the transmitter X100 e.
Transmitter M300
NOTE
Bridge terminal 2 and 3 in the EasyClean 150.
Set relay #1 (default) or #2, #3, #4 on the transmitter
to rinse contact.
Define rinse interval and rinse time according to the
process requirements on the transmitter M300.
Transmitter M700
NOTE
Bridge terminal 2 and 3 in the EasyClean 150.
Set relays K3 (or K1 or K2) to rinse contact and NO
(normally open). Define the rinse interval and rinse
time according to the process requirements.
Set input OK1 to HOLD (active 10–30 V) for the exter-
nal hold signal.
Transmitter Trb 8300
Define rinse interval and rinse time according to the
process requirements on the transmitter Trb 8300.
EasyClean 100 and 150 are designed to work together with METTLER TOLEDO transmitters.
Please consult relevant instruction manuals of third party transmitters for proper installation.
EasyClean 150
Wash
Hold
X100e
EasyClean 150
TB1(1/2 DIN)
TB2
or
M300
Hold
TB1B (1/4 DIN)
EasyClean 150
K3
OK1
M700
EasyClean 150
Wash
Hold
Trb 8300
EasyClean 100/150
52 500 178 D
23

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ingold easyclean 150

Inhaltsverzeichnis