Safety Instructions

Intended Use

Your weighing terminal is used for weighing. Use the balance exclusively for this purpose. Any other type of use and operation
beyond the limits of technical specifications without written consent from Mettler-Toledo, LLC is considered as not intended.
It is essential for the buyer to closely observe the installation information, product and system manuals, operating instructions and
other documentation and specifications. MT's warranty and any liability are expressly excluded for damages caused by non-
compliance with the applicable manuals.
Do not use the terminal in any environment or category other than those specified under Specifications.

Documentation

For further details on system configuration and operation, visit www.mt.com/IND500x. For product compliance information, visit
http://glo.mt.com/global/en/home/search/compliance.html.

Safety Warnings

DOWNLOAD and READ the Installation Guide BEFORE operating or servicing this equipment and FOLLOW all instructions carefully.
DO NOT INSTALL OR PERFORM ANY SERVICE ON THIS EQUIPMENT BEFORE THE AREA IN WHICH THE EQUIPMENT IS LOCATED HAS BEEN
SECURED AS NONHAZARDOUS BY PERSONNEL AUTHORIZED TO DO SO BY THE RESPONSIBLE PERSON AT THE CUSTOMER'S SITE.
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS
THAT MUST BE MADE WITH POWER ON. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN BODILY HARM AND/ OR PROPERTY
DAMAGE.
REFER TO THE DATA PLATE OF THE EQUIPMENT TO DETERMINE IF IT IS APPROVED FOR USE IN AN AREA CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES.
ALL EQUIPMENT MUST BE INSTALLED IN ACCORDANCE WITH THE INSTALLATION INSTRUCTIONS AND CONTROL DRAWING 30595335
RESPECTIVELY IN THE MANUFACTURER MANUALS OF THE ASSOCIATED APPARATUS. BE CAREFUL TO STUDY AND FOLLOW THE GUIDANCE
(GENERAL NOTES AND APPLICATION SPECIFIC NOTES) IN THE CONTROL DRAWING 30595335. DEVIATION FROM THE INSTRUCTIONS CAN
IMPAIR THE INTRINSIC SAFETY OF THE EQUIPMENT AND VOID THE AGENCY APPROVAL.
INSTALL CABLE SEAL BETWEEN DIFFERENTLY RATED AREAS PER COUNTRY-SPECIFIC REGULATIONS. FOR DETAILS, REFER TO CONTROL
DRAWING 30595335.
REFER TO THE DATA PLATE OF THE EQUIPMENT TO DETERMINE IF IT IS APPROVED FOR USE IN AN AREA CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES.
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD, CONNECT TO PROPERLY GROUNDED POWER SOURCE ONLY. DO NOT REMOVE THE
GROUNDING CONNECTION.
ENSURE PROPER EQUIPOTENTIAL GROUNDING OF THE EQUIPMENT, MOUNTING ACCESSORIES, AND THE SCALE BASE.
BE CERTAIN THAT THE COMMUNICATION CIRCUITS ARE WIRED EXACTLY AS SHOWN IN THE INSTALLATION SECTION OF ITS
CORRESPONDING USER MANUAL. IF THE WIRES ARE NOT CONNECTED CORRECTLY, THE EQUIPMENT OR INTERFACE BOARD MAY BE
DAMAGED.
THE MAINS CONNECTION OF THE POWER SUPPLY UNIT MUST BE MADE BY A PROFESSIONAL ELECTRICIAN AUTHORIZED BY THE OWNER
AND IN ACCORDANCE WITH THE RESPECTIVE TERMINAL DIAGRAM, THE ACCOMPANYING INSTALLATION INSTRUCTIONS AS WELL AS THE
COUNTRY-SPECIFIC REGULATIONS.
THE EXPLOSION-PROTECTED WEIGHING SYSTEM MUST BE CHECKED TO ENSURE COMPLIANCE WITH THE REQUIREMENTS FOR SAFETY
BEFORE BEING PUT INTO SERVICE FOR THE FIRST TIME, FOLLOWING ANY SERVICE WORK AND EVERY 3 YEARS, AT LEAST.
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS PRESENT.
AVOID DAMAGE TO THE SYSTEM COMPONENTS. IF DAMAGE OCCURS, IMMEDIATELY PUT THE SYSTEM OUT OF OPERATION. IMMEDIATELY
REPLACE DAMAGED SYSTEM COMPONENTS. REPAIRS MUST BE PERFORMED BY AUTHORIZED SERVICE PERSONNEL.
AVOID PLASTIC COVERS OVER THE EQUIPMENT. WEAR SUITABLE CLOTHING. AVOID NYLON, POLYESTER OR OTHER SYNTHETIC MATERIALS
THAT GENERATE AND HOLD CHARGE. USE CONDUCTIVE FOOTWEAR AND FLOORING.
IF THE KEYBOARD, DISPLAY LENS OR ENCLOSURE IS DAMAGED, THE DEFECTIVE COMPONENT MUST BE REPAIRED IMMEDIATELY. REMOVE
POWER IMMEDIATELY AND DO NOT REAPPLY POWER UNTIL THE DISPLAY LENS, KEYBOARD OR ENCLOSURE HAS BEEN REPAIRED OR
REPLACED BY QUALIFIED SERVICE PERSONNEL. FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.