Herunterladen Inhalt Inhalt Diese Seite drucken

Operation; Description - Endress+Hauser Memocheck Plus CYP01D Kurzanleitung

Referenzmittel zur anlagenqualifizierung und servicetool zur schnellen überprüfung von messsystemen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Memocheck Plus CYP01D / Memocheck CYP02D
9

Operation

9.1

Description

9.1.1 Memocheck Plus CYP01D
Memocheck Plus CYP01D is a set of 5 various sensor plug-in heads that provide data for the
simulation of pH values. Each plug-in head simulates the state of a corresponding sensor. The
following data are indicated and transmitted, they cannot be overwritten:
• Main measured value, e.g. pH
• Temperature
• Error message, e.g. glass breakage: One plug-in head transmits an error message.
The error message can be displayed or suppressed by switching the corresponding transmitter
function, e.g. glass SCS, on or off.
Transmitted measured values for pH glass electrodes (CYP01D-PG***)
Plug-in head
1
2
3
4
5
Standard data set of process data
All Memocheck Plus CYP01D plug-in heads contain the complete data set of a digital sensor.
This means they provide the same amount of information for display as digital sensors.
The stored data set cannot be overwritten which guarantees a safe use as qualification tool at
any time.
Among others, the following standard process data are stored in the plug-in heads:
• Total operating hours:
• Number of sterilisations:
• Max. temperature:
• Operating hours above 80 °C (176 °F):
• Operating hours above 100 °C (212 °F):
• Operating hours > 300 mV:
• Operating hours < -300 mV:
• First commissioning:
Endress+Hauser
pH value
0
4
7
10
14
Temperature
-10 °C (14 °F)
25 °C (77 °F)
60 °C (140°F)
90 °C (194 °F)
135 °C (275 °F)
8760 h
2
135 °C (275 °F)
14 h
9 h
15 h
0 h
01.06.2005 12:00
Operation
Glass SCS
o.k.
o.k.
o.k.
o.k.
Error
15