Mattress replacement side covers:
EN
1.
Remove all packaging.
Ensure bed side supports are extended
2.
and in the upright closed position.
3.
With the storage pocket facing outwards,
slip the cover over the side support.
4.
Repeat for the opposite side.
Note: If the side supports are to be used in the
open position, proceed as follows.
5.
Remove the additional support piece from
the storage pocket.
6.
Position the support piece in the gap
created by opening the side support.
Bring the support piece over the top of
the side support and locate between
the mattress and side support to form a
continuation of the mattress.
7.
To return the side supports to the upright
(closed) position, reverse Steps 5 and 6.
Do not use the mattress without a cover. It provides
a protective barrier against biological contamination
of the foam.

Technical Data

Pentaflex Plus 2-way mattress replacement system and side covers
Size (mm)
Product code
Material
Cover material
Patient weight:
Over 250 kg (551 lb / 39 stones)
Fire retardancy:
System meets the requirements of BS EN 597-1: 1995, BS EN 597-2: 1995,
and BS 6807: 2006, 5.
Cigarettes, lighters, naked flames and exposed heating elements e.g. in electric fires (and other
electrical appliances) are sources of fire hazard and could ignite bed linen, blankets and clothing. It is
4
CAUTION
Mattress replacement system
2060 X 920 X 172
450PM000/000
High resilience polyurethane foam
Polyurethane coated knitted fabric
essential that these fire sources are kept away from the mattress.
NB: It is highly dangerous to smoke in bed.
Decontamination
Mattress replacement system and side
covers:
Note: The side covers are a finished assembly
and must not be machine washed.
To Clean the Cover
Wipe all exposed surfaces with a cloth which
has been moistened in simple detergent and
water. Ensure to remove any organic debris
that may be adherent to the surfaces. Dry
thoroughly.
To Disinfect the Cover
For disinfection we recommend a chlorine-
releasing agent such as NaOCI or NaDCC
diluted to 1,000 ppm (range may vary
between 250 & 10,000 ppm depending on
contamination status and local policy).
Wipe with a cloth moistened with clean water
and dry thoroughly.
Alternatively, use a cloth moistened with
alcohol (70%) all allow to air dry.
Routine checks
The foam core and cover should be inspected
regularly for any signs of damage, soiling or
staining.
Side covers
1535 X 565 X 160
PMSR PBSR
High resilience polyurethane foam
Coated knitted fabric
FIRE WARNING