Disc Margin

Connecting RotoForce-1

Supplying Power

Electrical Connection

Connection to Compressed Air

RotoForce-1
Instruction Manual
Before final tightening of the two fastening screws or the bottom
nut you must check the margin between the specimen holder
plate hole and the edge of the grinding / polishing disc. This
distance should be:
Diameter of the Holes
40 mm
1½"
1¼"
30 mm
25 mm
Having adjusted the disc margin, tighten the two fastening screws or the
bottom nut firmly. The column should now remain in position.
RotoForce-1 is supplied with electric cable, air hose and
connection cable.
Check that the mains voltage corresponds to the voltage stated on the type
plate on the back of the machine.
The mains cable of RotoPol is either equipped with a standard
electric male plug, type "F" or with a male flat blade attachment
plug. If the type of plug delivered with the equipment has not
been approved for local use, the plug must be replaced with an
approved type. The leads have to be connected as following:
Yellow/green: earth
Brown: phase
Blue or black: neutral
n Insert the female plug of the power cable into the socket on
the back of the machine.
n Connect the male plug to a grounded electric outlet.
n Mount an air hose on the enclosed quick coupling and secure
with the enclosed hose clamp.
n Connect the quick coupling to the factory-mounted air hose of
RotoForce-1.
The air pressure must be between 5 bar (73 psi) and 6.5 bar (95 psi)
Disc Margin
1.5 mm
2.5 mm
6.0 mm
6.5 mm
9.0 mm
IMPORTANT