4.1 Cleaning the exhalation system

1. Dismantle the exhalation system
exhalation system", page
2. Clean the individual parts in accordance with the table below:
Action
Wash the individual parts in hot water and a mild detergent.
Clean individual parts thoroughly with a cloth or a soft brush
when washing.
Or: Put individual parts in the dishwasher (max. 65 °C).
3. Rinse the individual parts in clean water.
4. Allow the individual parts to air-dry.
5. Perform a visual inspection.
6. If necessary: Replace a damaged exhalation system.
7. Assemble the exhalation system
exhalation system", page
Discolorations do not impair the function of the exhalation
system.

4.2 Disinfecting the exhalation system

1. Clean individual parts
system", page
2. Immerse individual parts in a disinfectant solution.
|